Model Number N/A |
Device Problem
Tear, Rip or Hole in Device Packaging (2385)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/23/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that while preparing for the procedure, a hole was detected in both inner packaging.Another nail was used to complete the procedure.No adverse events have been reported as a result of the malfunction.Attempts have been made and additional information on the reported event is unavailable.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Confirmation through visual inspection of the returned product.The likely condition of the product when it left zimmer biomet was conforming to all specifications.The reported event could have occurred during shipping.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|