ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CONFIRM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM3500
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 08/13/2018
Event Type  malfunction  
Event Description
It was reported that during interrogation, prior to implant procedure, the implantable cardiac monitor displayed an error message while still in the box.The device was not implanted.
 
Manufacturer Narrative
Analysis of device image indicated that memory corruption occurred during the manufacturing process.Further testing was performed by reloading the product code and all tests results were normal, memory corruption could not be reproduced.A manufacturing anomaly may have occurred that resulted in the reported error message due to memory corruption.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7828397
MDR Text Key118626825
Report Number2017865-2018-12719
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 11/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/16/2019
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberS000055844
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/20/2018
Was the Report Sent to FDA? No
Date Manufacturer Received08/13/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2017865-10/11/19-001-C
Patient Sequence Number1
-
-