Related manufacturer reference numbers: 2017865-2018-12748, 2938836-2018-10101.
During an implant procedure, there was decreased r-wave sensing amplitude noted on the device; therefore, the physician attempted to disconnect the leads to remeasure the sensing value, but there was difficulty loosening the right atrial (ra) and right ventricular (rv) leads from the header of the device.
The device and both ra and rv leads were explanted and replaced to resolve this event.
The patient was stable.
|
The reported event of the doctor being unable to untighten the right atrial (ra)-setscrew was confirmed.
Electrical anomaly was not confirmed.
The analysis found on ra-setscrew was firmly stuck in its connector blocks and required destructive methods to separate them from the device connectors.
Analysis performed using scanning electron microscope confirmed the foreign material found in the connector block threads to be epoxy, which caused the setscrews to become lodged.
When the setscrew was tightened on the leads during the implant procedure the epoxy in the threads caused the setscrews to be locked in place which prevented removal of the leads.
As a result of these findings, abbott is performing further investigation.
|