• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B.BRAUN SURGICAL SA MONOPLUS VIOLET 5/0 (1) 45CM DSMP13 (M); SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B.BRAUN SURGICAL SA MONOPLUS VIOLET 5/0 (1) 45CM DSMP13 (M); SUTURE Back to Search Results
Model Number C0024916
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/30/2018
Event Type  malfunction  
Event Description
Country of complaint: (b)(6).Needle detachment.
 
Manufacturer Narrative
Samples received: none.Analysis and results: there are no previous complaints of this code batch of which we manufactured and distributed (b)(4) in the market.There are no units in stock in b.Braun surgical's warehouse.As no samples have been received and no units are available in b.Braun surgical, s.A.We have only reviewed the batch manufacturing record and this product had a normal process and the results during the process fulfill usp/ep and b.Braun surgical requirements.Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done and the case is not confirmed due to lack of evidence.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyse it.Please note that when no samples are received our analysis is very limited.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONOPLUS VIOLET 5/0 (1) 45CM DSMP13 (M)
Type of Device
SUTURE
Manufacturer (Section D)
B.BRAUN SURGICAL SA
121 carretera de terrassa
rubi, 08191
SP  08191
MDR Report Key7834245
MDR Text Key118885451
Report Number3003639970-2018-00512
Device Sequence Number1
Product Code NEW
Combination Product (y/n)N
PMA/PMN Number
K031216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC0024916
Device Catalogue NumberC0024916
Device Lot Number118065
Was Device Available for Evaluation? No
Distributor Facility Aware Date08/13/2018
Date Manufacturer Received09/20/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-