The production device history record (dhr) for this iabp unit is not required to be reviewed per sop (b)(4) since the device manufacture date is greater than one year from the event date.
A getinge field service engineer (fse) was dispatched to evaluate this unit.
The fse tried to switch the machine on but he got the same reported failures.
There was no power-up failure codes on the monitor because of booting not completed.
But on the other hand, the system diagnostics mode has booted normally.
The fse was trying to switch on/off the machine a few times more but machine was not able to started operate in operate mode.
The fse took the console out from the cart, took the cover from console out and saw on the executive processor board led readouts ¿f0 ¿.
He tried off/on thy system again and it was successful.
The system booted correctly up with the power-up failures: power-up test fails code #10, the transducers not calibrated and no backup battery detected.
As per the manual, the fse performed the 30 psi transducers and the calibration, but the problem was not fixed.
The fse performed 30 psi transducers and calibration again.
Transducer gain checks were done according to the service manual and all passed.
The fse also found the date, time and user language have been set incorrectly, and he fixed the issue and set up the correct values.
After off/on the cardiosave system the black touch screen has appeared again with the continuous tone.
All manifold test was done with a pass result.
Front end board, executive processor board and back plane were not damaged by saline.
The other device was used instead.
|