MAUDE Adverse Event Report: DATASCOPE MAHWAH CS100 SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE MAHWAH CS100 SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number N/A
Device Problem No Device Output (1435)
Patient Problem No Patient Involvement (2645)
Event Date 08/15/2018
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this iabp unit was reviewed and no non-conformances related to the reported event were noted. A distributor's field service engineer (fse) evaluated this unit and was able to reproduce the reported failure. To fix the issue, the fse replaced the display. The cardiosave unit then was running normal.
 
Event Description
It was reported that when the customer power on the cs100 intra-aortic balloon pump (iabp), he found that the screen turned blue and there was no alarm sound. After repeated reboots, the fault still cannot be eliminated. They stopped using this iabp. There was no patient involvement and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCS100
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7847777
MDR Text Key119504649
Report Number2249723-2018-01523
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K031636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 09/05/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberN/A
Device Catalogue Number0998-00-3013-XX
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Device Age YR
Event Location No Information
Date Manufacturer Received08/15/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/16/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-