BOSTON SCIENTIFIC - COSTA RICA (COYOL) OPTICROSS¿; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number H749518080120
Device Problem Poor Quality Image (1408)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/03/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: the device was returned for analysis.Device analysis revealed a kink in the sheath assembly from femoral marker to the proximal end and a kink in the sheath assembly from femoral marker to the distal end.A damage observed at the guidwire exit hole port, additionally, a damage was encountered in the imaging window (it was observed that the imaging core protruded from the damaged imaging window).The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
Reportable based on device analysis completed on 16-aug-2018.It was reported that lost image occurred.An opticross imaging catheter was selected for use.During the procedure, the physician inserted the opticross imaging catheter into the blood vessel as the stage of the procedure.Lost image occurred during manual imaging.No patient complications were reported.However, returned device analysis revealed an imaging core protruded from the damaged imaging window.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS¿
Type of Device
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7847979
MDR Text Key119647907
Report Number2134265-2018-07412
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/25/2019
Device Model NumberH749518080120
Device Catalogue Number51808-012
Device Lot Number22171724
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/16/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/25/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-