Results: the pump max was opened by the penumbra investigator and corrosion was observed on the cylinder piston crowns.Conclusions: evaluation of the returned pump max confirmed the device was unable to produce sufficient vacuum pressure.The pump housing was removed and the vacuum pump was opened by a penumbra investigator.It was observed that the piston crown was corroded.The observed corrosion likely contributed to the insufficient vacuum pressure.The root cause of the corrosion could not be determined.Penumbra pumps are visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 110 (pump max).During the procedure, the hospital staff noticed that the pump max was unable to produce maximum aspiration; the pump max was able to reach only -20 inhg.However, the procedure was able to be completed using the pump max.There was no report of an adverse effect to the patient.
|