• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ZNN TEAR DROP GUIDE WIRE 3.0 MM DIAMETER 70 CM LENGTH ROD, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ZNN TEAR DROP GUIDE WIRE 3.0 MM DIAMETER 70 CM LENGTH ROD, FIXATION Back to Search Results
Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Patient Involvement (2645)
Event Date 08/09/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4). Customer has indicated that the product is in process of being returned to zimmer biomet for investigation. Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It is reported that upon receipt at the distributorship, there was a stain found on the product. There was no adverse events reported as a result of this malfunction as there was no patient involvement.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information. Updated: reported stains were confirmed by visual examination. It was determined the stain was caused by the way the tear drop guide wires are fixtured during passivation. The guide wires are placed on a rack with two or more contact points. These contact points cause a small portion of the guide wire to not receive a full passivation coating, leading to an inconsistent coating appearing as a "stain" or mark. These stains are not corrosion, and do not impact the fit, form, or function of the guidewires; they are merely cosmetic. The product is conforming per zimmer biomet's criteria. No product problem has been identified and the devices conform to their specifications. If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly. Zimmer biomet will continue to monitor for trends.
 
Event Description
No further information available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameZNN TEAR DROP GUIDE WIRE 3.0 MM DIAMETER 70 CM LENGTH
Type of DeviceROD, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7850345
MDR Text Key119344321
Report Number0001822565-2018-04827
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK101622
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial,Followup
Report Date 12/05/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberN/A
Device Catalogue Number47249009800
Device Lot Number64070834
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2018
Is the Reporter a Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received12/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A

-
-