SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616
Device Problems Overheating of Device (1437); Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/27/2018
Event Type  malfunction  
Event Description
It was reported that motor died and then got warm.A backup device was available to complete the procedure with no delay or patient injuries.
 
Manufacturer Narrative
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A relationship, if any, between the subject device and the reported event could not be determined.No containment or corrective actions are recommended at this time.A review of the device history record shows there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Manufacturer Narrative
H3, h6: the reported device, used in treatment, was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Product failed functional testing with blade stall error and overheating.Cause of overheating and errors is a corroded motor/gearbox.The housing and the gearbox were removed from the motor.The gearbox was found to be jammed.The complaint was confirmed and the root cause has been determined to be corrosion of the motor and gearbox assembly.A blade stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing.Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key7858535
MDR Text Key119798659
Report Number1643264-2018-00684
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup,Followup
Report Date 07/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2020
Date Manufacturer Received06/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-