(b)(6).
Device evaluation: the failure was confirmed by a company representative during the testing.
The device was evaluated and it was determined that the membrane was wrapped incorrectly by the operator during the tight wrapping operation.
This is evident by the folds present in the membrane after it was pulled from retainer, see.
When comparing the improperly folded membrane in, to a properly folded membrane also tested in this protocol it is clear that the membrane was wrapped incorrectly.
An additional comparison can also be made between the improperly folded membrane and the non-acceptable sample pictures from the linear 7.
5fr.
Iab tight wrap packaging & inspection procedure.
(b)(4).
|