• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. ANESTHESIA SUPPORT TRAY; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. ANESTHESIA SUPPORT TRAY; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number 332114
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device involved has not been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event Description
As reported by the user facility: end user stated that her staff is noticing either increased moisture or excess lubricant on the inside of the syringes.It was stated that it seems like its lubricant.No injury reported.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Multiple unsuccessful attempts were made to obtain a sample.Without the actual device, a thorough investigation could not be performed.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If a sample and/or any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANESTHESIA SUPPORT TRAY
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
MDR Report Key7859870
MDR Text Key119932436
Report Number2523676-2018-00074
Device Sequence Number1
Product Code CAZ
UDI-Device Identifier04046964177507
UDI-Public04046964177507
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 10/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2019
Device Catalogue Number332114
Device Lot Number0061624161
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/29/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-