Catalog Number 72200616
Device Problem Failure to Shut Off (2939)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/17/2018
Event Type  Malfunction  
Event Description

It was reported that the device kept running unintendedly. Event occurred before procedure. No patient injuries reported.

 
Manufacturer Narrative

The device was received for evaluation. There was no relationship found between the returned device and the reported incident. A visual inspection was performed and no issues were noted. Complaint of mdu running by itself could not be reproduced. Mdu failed functional testing with hand piece sensor fault. Cause of hand piece sensor fault is a defective hall board. The complaint of running by itself was not confirmed but the root cause of sensor fault has been determined to be a defective electronic component on the hall board. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key7860822
MDR Text Key119818909
Report Number1643264-2018-00682
Device Sequence Number1
Product Code HAB
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Type of Report Initial,Followup
Report Date 10/10/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received09/10/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue Number72200616
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer08/29/2018
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/09/2018
Was Device Evaluated By Manufacturer? Yes
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-