MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 9735665
Device Problems Erratic or Intermittent Display (1182); No Display/Image (1183)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/15/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site. A medtronic representative went to the site to test the equipment. Testing revealed that the monitor was tested by moving in all directions and the cable connections to and from the monitor were checked. External hdmi cable was wiggled and there were no issues with power to the monitor. The reported issue could not be replicated. The system then passed the system checkout and was found to be fully functional.
 
Event Description
Medtronic received information regarding a navigation system being used for a posterior lumbar interbody fusion (plif). It was reported that there was intermittent display on the main cart monitor. It was reported that pre-operatively, the main cart was moved and the video input went out. The camera cart monitor was used to finish the case without any issues. There was no reported patient impact and no reported surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION S8 PREMIUM SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
prashanth gali
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7861282
MDR Text Key119791192
Report Number1723170-2018-04497
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation
Type of Report Initial
Report Date 09/10/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number9735665
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/22/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-