• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 53401
Device Problem Computer Software Problem (1112)
Patient Problem No Patient Involvement (2645)
Event Date 08/28/2018
Event Type  malfunction  
Manufacturer Narrative
Manufacturer's analysis indicated that the external pulse generator (epg) experienced a system error and had a broken cover.The epg was repaired and returned to service.This device was reported as included in the field action noted and returned product investigation found the device performed as described in the field action.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The external pulse generator (epg) was returned for field action and subsequently tested out of specification during manufacturer's analysis.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7862046
MDR Text Key119791265
Report Number3004593495-2018-00856
Device Sequence Number1
Product Code DTE
UDI-Device Identifier00643169723160
UDI-Public00643169723160
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date 09/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number53401
Device Catalogue Number53401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2018
Date Manufacturer Received09/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2317-2018
Patient Sequence Number1
-
-