Catalog Number 9735669
Device Problem Device Difficult to Setup or Prepare (1487)
Patient Problem No Patient Involvement (2645)
Event Date 08/17/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Unique device identification (udi) not available.\ device manufacturing date is dependent on lot number/serial number, therefore, unavailable.A software investigation analysis was initiated to determine the probable cause of the issue through exam analysis.Analysis found that when exams were loaded onto navigation system, one exam was valid and the other two would not load due to only containing 2-3 slices.
 
Event Description
Medtronic received information regarding a navigation system.It was reported outside of a procedure when a technician was testing the new system with a demo exam, one exam failed to load.There was no patient present.
 
Manufacturer Narrative
Udi provided.Updated to proper value.Device manufacture date provided.Updated to proper value.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 EM ENT SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
koko tyler
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208903200
MDR Report Key7863951
MDR Text Key119804942
Report Number1723170-2018-04521
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9735669
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/26/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-