The production device history record (dhr) for this intra-aortic balloon pump (iabp) unit is not required to be reviewed as per the company standard operating procedure since the device manufacture date is greater than one year from the event date.
During a pm service, the getinge field service engineer (fse) evaluated the front-end board and noticed it would not calibrate.
To fix the issue, the fse replaced front end board as per manufacturer specifications.
The fse completed the pm and the unit passed all functional and safety tests per factory specifications.
The iabp was then released to the customer for return to clinical service.
The full name of the event site noted (b)(6) hospital.
|