Model Number PM3262
Device Problems Signal Artifact/Noise (1036); No Pacing (3268)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/12/2018
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.  based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
During follow-up, noise was observed on the device resulting in pacing inhibition. Noise was replicated with provocative testing. The patient has a non-abbott right ventricular lead and it is unknown if the lead or device was causing the inhibition of pacing. Further information was requested but not received. The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameQUADRA ALLURE MP RF CRT-P
Type of DeviceIMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7884803
MDR Text Key120532904
Report Number2017865-2018-14003
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 09/18/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date11/30/2017
Device Model NumberPM3262
Device Lot NumberP000029354
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received08/30/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/17/2016
Is the Device Single Use? No Answer Provided
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-