• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S8 PREMIUM SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9735665
Device Problems Communication or Transmission Problem (2896); Data Problem (3196)
Patient Problem No Patient Involvement (2645)
Event Date 09/04/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.No additional information has been provided/submitted to the manufacturer for an evaluation to be conducted.Device not returned.
 
Event Description
Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that the fmri scans loaded from the electronic picture archiving and communication systems (pacs) showed up as 0 slices.It was unclear if the images had been exported as grey scale or colorized.It was also noted that the scan thumbnail images were black.A preprocessed fmri with brainlab was run, then an attempt was made to send the deconstructed scans in greyscale to the navigation system.Ten slices loaded which did not meet the minimum number of slices required for navigation.There was no patient present when this issue was observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S8 PREMIUM SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
erika mitchellette
navigation customer quality
826 coal creek circle
louisville, CO 80027
MDR Report Key7891040
MDR Text Key120832613
Report Number1723170-2018-04853
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735665
Device Catalogue Number9735665
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/19/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-