DEPUY SYNTHES PRODUCTS LLC RECON SAGITTAL SAW WITH KEY FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 05.001.240 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
Not Applicable (3189)
|
Event Date 09/05/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the handle of the sagittal saw with key device was blocked.It was noted that the button on the t-handle was blocked.It was reported that there was a ten-minute delay in the procedure due to the event, as an unspecified spare device was available for use.It was reported that the procedure was successfully completed.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.The device was evaluated and it was determined that the trigger was blocked, jammed, and moving heavily.It was noted that the trigger was defective, and the trigger unblocks.It was further determined that the device failed pretest for check proper function of the trigger.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to premature wear.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|