DEPUY SYNTHES PRODUCTS LLC RECON SAGITTAL SAW WITH KEY FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 05.001.240
Device Problem Mechanical Jam (2983)
Patient Problem Not Applicable (3189)
Event Date 09/05/2018
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the handle of the sagittal saw with key device was blocked.It was noted that the button on the t-handle was blocked.It was reported that there was a ten-minute delay in the procedure due to the event, as an unspecified spare device was available for use.It was reported that the procedure was successfully completed.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.The device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.The device was evaluated and it was determined that the trigger was blocked, jammed, and moving heavily.It was noted that the trigger was defective, and the trigger unblocks.It was further determined that the device failed pretest for check proper function of the trigger.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to premature wear.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECON SAGITTAL SAW WITH KEY FOR TRS
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7893020
MDR Text Key120830423
Report Number8030965-2018-56455
Device Sequence Number1
Product Code GEY
UDI-Device Identifier7611819462588
UDI-Public(01)7611819462588(11)160406
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.240
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2018
Date Manufacturer Received11/30/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-