Model Number 9770501
Device Problem Product Quality Problem (1506)
Patient Problems Hemorrhage/Bleeding (1888); Perforation of Vessels (2135)
Event Type  Injury  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer for evaluation.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
 
Event Description
It was reported by the vat rn that the needle visualization on the site rite 8 is poor.The rn uses the system for training on picc, accucath and midlines and finds it very difficult to train when the needle echo is not present when used on a vein block or patient.As a result of the poor needle visualization the rn's in training have been documented to hit arteries, nerves, caused vessels to bleed yet site rite 8 indicated a successful vascular access obtained in about 50% of the attempts.The rn stated image quality is fine for most end users and he can visualize the vascular structure and other shallow anatomical structures but when free-handing a catheter placement and looking for the indentation of the vascular structure puts the needle too close to the structure to avoid impact or injury to the targeted structure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
DYMAX CORP.
141 zehner school road
zelienople PA 16063
Manufacturer Contact
shelly gilbert
605 n. 5600 w.
salt lake city, UT 84116
8015225640
MDR Report Key7899675
MDR Text Key121436719
Report Number3006260740-2018-02544
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741098338
UDI-Public(01)00801741098338
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Nurse
Type of Report Initial
Report Date 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number9770501
Device Catalogue Number9770501
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Hospital
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-