Model Number 9770501 |
Device Problem
Product Quality Problem (1506)
|
Patient Problems
Hemorrhage/Bleeding (1888); Perforation of Vessels (2135)
|
Event Type
Injury
|
Manufacturer Narrative
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer for evaluation.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
|
|
Event Description
|
It was reported by the vat rn that the needle visualization on the site rite 8 is poor.The rn uses the system for training on picc, accucath and midlines and finds it very difficult to train when the needle echo is not present when used on a vein block or patient.As a result of the poor needle visualization the rn's in training have been documented to hit arteries, nerves, caused vessels to bleed yet site rite 8 indicated a successful vascular access obtained in about 50% of the attempts.The rn stated image quality is fine for most end users and he can visualize the vascular structure and other shallow anatomical structures but when free-handing a catheter placement and looking for the indentation of the vascular structure puts the needle too close to the structure to avoid impact or injury to the targeted structure.
|
|