The 3500cp-c mixer was evaluated and the reported issue of the failure to alarm when the air input was disconnected was unverifiable.The reported issue could not be replicated during evaluation testing.A search of device records indicates no history of service since the unit's manufacturing in (b)(6) 1992.This period of time without maintenance is not advised, as the service life for a 3500 series mixer is 2 years, at which point the sechrist mixer must be thoroughly overhauled by a sechrist trained and approved service technician.This is in accordance with the 3500 series air/oxygen mixer (b)(4).
|
It was reported that during clinical use when the mixer was placed on bypass, it was noticed that the patient showed no o2 saturation and the mixer did not alarm.Procedure was quickly stopped and completed with another mixer.No patient injury was reported.Reportedly, there was not an oxygen analyzer and alarm in-line downstream from the mixer.
|