Model Number MONITOR POWER CORD
Device Problem No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/27/2018
Event Type  malfunction  
Manufacturer Narrative
The device has not been evaluated yet for investigation purpose. Once the evaluation will be performed, a follow-up medwatch report will be submitted.
 
Event Description
The surgeon stated there was no video signal on the monitor, and a nurse stated there was a signal before. The surgeon restarted the device and the video signal was there again. A field service engineer went on customer site and found that the isolation of the power cord of the monitor was damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameROSA SURGICAL DEVICE
Type of DeviceCOMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR 34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR 34000
Manufacturer Contact
jay sharma
1520 tradeport drive
jacksonville, FL 32218
9047414400
MDR Report Key7901619
MDR Text Key121467422
Report Number3009185973-2018-00248
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
PK101791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/06/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberMONITOR POWER CORD
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received02/21/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A

-
-