Model Number 809810
Device Problem Insufficient Heating (1287)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/30/2018
Event Type  malfunction  
Event Description
It was reported that during use of the device for a cardiopulmonary bypass (cpb) procedure, the heater-cooler displayed an "e08" error message on channel a.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.Per clinical review: the heater-cooler had an incident during a cpb procedure with the patient cooling/warming side.The heater-cooler was set up and worked for the early part of the cpb procedure on (b)(6) 2018.During rewarming of the patient, the perfusionist had the heater-cooler give an "e08" error message on the patient side channel.The perfusionist had to the exchange the unit out, for he was unable to warm the patient with the e08 failure.The perfusionist proceeded accordingly with the working unit.The incident did not delay the surgical procedure, for the perfusionist was able to exchange the unit in a timely manner to give the surgical team time to perform the repair.There was no blood loss or harm associated with the event.
 
Manufacturer Narrative
The field service representative (fsr) verified the reported complaint.He replaced the heater assembly on channel a.The unit operated to the manufacturer's specifications.The suspect part was returned to the manufacturer for further evaluation.
 
Manufacturer Narrative
The reported complaint was confirmed.During laboratory analysis, the product surveillance technician (pst) measured the resistance of the heater elements and found the heater element supplied by the blue wires to be open.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEM
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-HX2 TEMPERATURE MANAGEMENT
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7903393
MDR Text Key122155643
Report Number1828100-2018-00492
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K071521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number809810
Device Catalogue Number809810
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2018
Was the Report Sent to FDA? No
Date Manufacturer Received11/21/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-