• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT; BZE Back to Search Results
Model Number RT266
Device Problems Leak/Splash (1354); Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 08/27/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint rt266 infant dual heated evaqua2 breathing circuit is currently en route to fisher & paykel healthcaare in (b)(4) for investigation.We will provide a follow-up report upon completion of our investigation.
 
Event Description
A hospital in (b)(6) reported via a distributor that a leak error occurred with the rt266 infant dual-heated evaqua2 breathing circuit prior to patient use.There was no patient involvement.
 
Event Description
A hospital in korea reported via a distributor that a leak error occurred with the rt266 infant dual-heated evaqua2 breathing circuit prior to patient use.There was no patient involvement.
 
Manufacturer Narrative
(b)(4).Method: the complaint rt266 infant dual heated evaqua2 breathing circuit was returned to fisher & paykel healthcare in new zealand where it was visually inspected and pressure tested for leak.Results: visual inspection revealed no damage to the breathing circuit and dryline.The pressure test revealed that the subject breathing circuit was within specification.Conclusion: we were unable to determine what may have caused the leak as reported by the customer as no fault was found with the returned device.However it is possible that the leak experienced by the customer was caused by a loose connection in their setup.All rt266 infant dual-heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.Our user instructions that accompany the rt266 infant dual heated evaqua2 breathing circuit state the following: check all connections are tight before use.Perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Set appropriate ventilator alarms.Do not soak, wash, sterilize or reuse this product.Avoid contact with chemicals, cleaning agents or hand sanitizers.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
MDR Report Key7908600
MDR Text Key121858598
Report Number9611451-2018-00816
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
PMA/PMN Number
K103767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,user faci
Type of Report Initial,Followup
Report Date 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT266
Device Catalogue NumberRT266
Device Lot Number2100429630
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2018
Date Manufacturer Received08/27/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-