MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. ANTHEM RF CRT-P IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. ANTHEM RF CRT-P IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3212
Device Problem Premature End-of-Life Indicator (1480)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/19/2018
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.  based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
During follow-up, the device was interrogated and an end of life (eol) status was displayed. The patient had received a cardioversion prior to interrogation, which the physician believes is the reason for the eol trigger. The device was restored to normal function and interrogated, an eol status was displayed a second time. Device explant is anticipated and the patient was in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameANTHEM RF CRT-P
Type of DeviceIMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7910001
MDR Text Key124681591
Report Number2017865-2018-14379
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/26/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date12/31/2012
Device Model NumberPM3212
Device Lot Number3434038
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received10/16/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/28/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 09/26/2018 Patient Sequence Number: 1
-
-