• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GUIDEZILLA II CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GUIDEZILLA II CATHETER, PERCUTANEOUS Back to Search Results
Model Number 1871
Device Problem Device Damaged by Another Device (2915)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/05/2018
Event Type  malfunction  
Event Description
It was reported that the stent caught on the guidezilla and got deformed. The target lesion was located in a calcified vessel. A guidezilla ii guide extension catheter and a 4. 00 x38 synergy drug eluting stent were used in performing a percutaneous coronary intervention. During procedure, the synergy stent did not cross the lesion when advanced through the guidezilla. Consequently, the stent delivery system was brought back into the guidezilla but the distal end of the stent caught on the distal end of the guidezilla; deforming the stent. The stent delivery system and guidezilla were then removed together from the patient's body and were completely intact. The procedure was completed with a new guidezilla and a new stent. No patient complications were reported and patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameGUIDEZILLA II
Type of DeviceCATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key7915643
MDR Text Key122013561
Report Number2134265-2018-61347
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163314
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial
Report Date 09/27/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/08/2020
Device Model Number1871
Device Catalogue Number1871
Device Lot Number0022352549
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/06/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/10/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Initial

-
-