(b)(4).Device evaluation by manufacturer: a field service engineer (fse) was at customer's site to resolve reported issue.Fse inspected instrument and found damaged seal on filter fitting.Fse replaced the line filter tubing due to damaged seal at filter fitting.Fse restarted analyzer, checked pressure and tested for leaks.Fse also ran several samples to confirm analyzer is operating as intended.Customer successfully completed quality controls and sample precision.The g8 analyzer is functioning as expected.No further action was required by the fse.A 13-month complaint history review and service history review for similar complaints was performed for serial (b)(4).There were no similar complaints found during this time.The g8 operator's manual under chapter 3, assay operations, section 7, states the following: 7.Other items to check: check the flow line connections, particularly the filter and the column inlet and outlet for leaks during warming up operations.Tighten the connection if a leak is found.The most probable cause of the reported event was due to damaged line filter tubing seal.
|
A customer reported liquid leak and low pressure on the g8 analyzer.Customer states that after changing the pre-filter, liquid was leaking from the pre-filter housing and the tubing going into the housing was loose.Technical support specialist instructed the customer to unscrew fitting and try to retighten the fitting.Customer stated the tubing is still loose and pulls out.A field service engineer (fse) was dispatched to address the reported event, which could have resulted in delay of results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|