Model Number 3CX*FX25RWC |
Device Problem
Partial Blockage (1065)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/03/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion.(b)(4).Results: results pending completion of evaluation.Conclusions: conclusion not yet available-evaluation in progress.
|
|
Event Description
|
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, it was impossible to turn at the nominal flow rate while removing the bubble.Tubing was twist, crushed under the protection spiral located at the oxygenator inlet, still impossible to turn the flow rate.No known impact or consequence to patient.The product was not changed out.It is unknown if there was a delay in the procedure, or if the surgery was completed successfully.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on october 1, 2018.(b)(4).All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|
Event Description
|
New information received indicating that by cutting the tube and placing it back in the circuit resolved the issue.There was no delay on the procedure.The procedure was completed successfully.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations.(b)(4).The sample was not returned; therefore no investigation could be completed.Additionally, the complaint is not related to the fx oxygenator, and it is a tubing issue in terumo uk produced tubing packs.The most likely cause of the low flow rate is compromised tubing after removing the bubble.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|