RESPIRONICS CALIFORNIA, INC V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Blower (738); Temperature Problem (3022)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
Malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2018.
Date of report: 2oct2018.
Customer reports single instance of high blower temp, code 1122.
Customer reported ran additional tests and no errors occurred.
Customer will continue to monitor unit until annual performance maintenance.
Issue was resolved and returned to clinical use.
No parts were returned for failure investigation; therefore, the root cause at the component level could not be determined.
|
|
Event Description
|
Customer reports single instance of high blower temp, (error code 1122).
No patient/user harm reported.
The event date was not specified; estimate used.
|
|
Search Alerts/Recalls
|
|
|