• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SHUNT SENSOR SYS500 BLOOD GAS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SHUNT SENSOR SYS500 BLOOD GAS MONITOR Back to Search Results
Model Number CDI510H
Device Problem Leak/Splash (1354)
Patient Problem Blood Loss (2597)
Event Date 09/13/2018
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed. Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available. (b)(4).
 
Event Description
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, a blood leakage was observed while on extracorporeal circulation. There was a blood loss of 10 ml or less. Product was changed out. Procedure was completed successfully.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on october 2, 2018.   (b)(4). All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations. (b)(4). The returned sample was visually inspected and was noted to contain dried blood within the threads of the large bore cap. It was leak tested (as received) by connecting with the calibrated manometer, submerged into a water bath, pressurized up to 1030 mmhg, and no leaks was noted. The large blue cap was attempted to be loosened; however, it was not able to be loosened due to the dried blood remained within the threads. The unit was left to soak in a di water for 24 hours. The large bore adapter blue cap was then able to loosened and retightened by hand, and leak tested for the second time. It was submerged into a water bath, pressurized up to 1030 mmhg. No leaks was noted. A retention sample from the same lot number was visually inspected and confirmed to have no traces of buffer on the outside of the unit or inside the pouch. It was then pressurized with air up to 1030 mmhg, submerged in a water bath, and no leaks were noted. The root cause for this event was determined to be the large blue vent cap for shunt sensor was not fully tightened either during setup of the circuit, or after the gas calibration. When the large blue vent cap was loosened, it had not been re-tightened fully prior to use in the line, causing a leak from the cap. All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSHUNT SENSOR SYS500
Type of DeviceBLOOD GAS MONITOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key7928178
MDR Text Key122458437
Report Number1124841-2018-00246
Device Sequence Number1
Product Code DRY
Combination Product (y/n)N
PMA/PMN Number
K972962
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 10/31/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date04/30/2019
Device Model NumberCDI510H
Device Catalogue NumberN/A
Device Lot NumberWF14E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2018
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received10/29/2018
Is This a Reprocessed and Reused Single-Use Device? No

-
-