Model Number R SERIES
Device Problem Sparking (2595)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that during biomed testing, the device's main plug began to spark once it was plugged in. Complainant indicated that there was no patient involvement in the reported malfunction.
 
Manufacturer Narrative
Zoll medical uk evaluated the device and the reported malfunction was not replicated or confirmed. The device was put through extensive testing which included bench handling and functional testing without duplicating the malfunction. The customer's pigtail and ac cord was observed to have burns and was replaced, unable to determined how burn damage occurred. The device was recertified and returned to the customer. No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameR SERIES DEFIBRILLATOR
Type of DeviceDEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key7929617
MDR Text Key122475768
Report Number1220908-2018-02820
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeGB
PMA/PMN Number
K060559/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation
Type of Report Initial,Followup
Report Date 09/12/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received09/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-