• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. EVERFLO OXYGEN CONCENTRATOR; GENERATOR, OXYGEN, PORTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. EVERFLO OXYGEN CONCENTRATOR; GENERATOR, OXYGEN, PORTABLE Back to Search Results
Model Number 1020020
Device Problem Insufficient Flow or Under Infusion (2182)
Patient Problem Death (1802)
Event Date 09/27/2018
Event Type  Death  
Event Description
The manufacturer received information alleging a patient expired while using an everflo oxygen concentrator with a ventilator.The investigation is still ongoing.A follow up report will be submitted when the manufacturer has completed the investigation.
 
Manufacturer Narrative
The manufacturer previously reported a patient allegedly expired while using an everflo oxygen concentrator with a ventilator.The device was returned to a third party service center for evaluation.The device was tested and was found to operate and alarm to design specifications.The manufacture concludes the device did not cause or contribute to the patient's death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLO OXYGEN CONCENTRATOR
Type of Device
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
175 chastian meadows court
kennesaw GA 30144 3724
Manufacturer (Section G)
RESPIRONICS, INC.
175 chastian meadows court
kennesaw GA 30144 3724
Manufacturer Contact
adam price
312 alvin drive
new kensington, PA 15068
7243349303
MDR Report Key7929997
MDR Text Key122497642
Report Number1040777-2018-00040
Device Sequence Number1
Product Code CAW
UDI-Device Identifier00606959032774
UDI-Public00606959032774
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K061261
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1020020
Device Catalogue Number1020020
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/11/2018
Date Manufacturer Received10/11/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/18/2017
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age84 YR
-
-