Catalog Number 199721000S
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/05/2018
Event Type  malfunction  
Manufacturer Narrative
Product complaint #
=
> (b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The following was reported: product code : 199721000s.Lot/batch/exp: 169336.Did the event happen during a procedure? yes were you in the procedure at the time of the event? yes.Was the product being used in a clinical trial? no.Event outcome/how was it managed? a unitized set screw was used with the reduction tool and final tightening was achieved.Was there a patient impact or was the procedure extended greater than 30 minutes due to the failure? no.Has the reporter facility indicated there may be legal action? no.Is the product available for return? if the item is returned from cssd following decom please give a detailed explanation of the event: mr (b)(6) was attempting final tightening and on attempting to complete final tightening of the outer segment of the dual innie with the castle tightener the thread of the nut stripped entirely.A quick stick was also engaged at this stage.A unitised set screw was used instead and final tightening was achieved.
 
Manufacturer Narrative
Product complaint #: (b)(4).Udi: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Visual examination found that the threads on the set screw were torn off completely.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.With the information provided, a definitive root cause for the torn threads on the 5.5 expedium verse di set screw cannot be determined.Noted damage suggests that inadvertently cross threading of the setscrew occurred upon insertion into the tulip head.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 EXP VERSE DI SET SCR
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key7931188
MDR Text Key124813544
Report Number1526439-2018-50941
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034466125
UDI-Public(01)10705034466125
Combination Product (y/n)N
PMA/PMN Number
K142185
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2022
Device Catalogue Number199721000S
Device Lot Number169336
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/21/2018
Date Manufacturer Received10/08/2018
Is This a Reprocessed and Reused Single-Use Device? Yes
Patient Sequence Number1
-
-