LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problems Melted (1385); Loss of Power (1475)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/08/2018
Event Type  Malfunction  
Manufacturer Narrative

The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601). Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). A livanova field service representative was dispatched to the facility to investigate. The service representative was able to confirm the reported issue and traced the failure to a defective cable. The technician replaced the cable to resolve the reported failure. A burned through cable could be identified as root cause for the reported issue however it could not be determined why the cable got burned through. A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.

 
Event Description

Livanova (b)(4) received a report that a heater-cooler system 3t lost power during procedure. In addition a burning smell could be detected when the device turned off. There was no report of patient injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7933677
MDR Text Key122641101
Report Number9611109-2018-01273
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Remedial Action Other
Type of Report Initial
Report Date 10/04/2018
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/04/2018
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number16-02-80
Was Device Available For Evaluation? No Answer Provided
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received09/08/2018
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured12/11/2001
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-