On september 4th and 5th of 2018, the user facility reported the following to richard wolf medical instruments (rwmic): "during a case, the suction was not powerful enough to proceed.They switched out the pumps and everything was fine." will the device be returned? yes.Was the device being used during a procedure when the issue occurred? yes.Specifically, was the device being used on a patient when the issue occurred? yes.Was there any injury or illness to patient or other personnel due to issue? no.Did the issue cause a delay in the procedure being performed that put the patient at risk? yes.Was there a similar back-up device available for use? yes.Was the scheduled procedure completed? yes.How was the patient anesthetized? yes.
|
Follow up report #1 is to provide fda with new and changed information: the following fields contain new or changed information: a3, b5, f3-5f7, f8, f9, f11, f13,g4, g7, h2, h3,h4, h6, h10.Richard wolf medical instruments corporation (rwmic) considers this case closed.In the event that rwmic receives additional information a follow up report will be submitted to the fda.Caution! for therapeutical use it is necessary to keep an adequate replacement unit at hand should the device fail.Important! run through the checks before and after each use.Do not use products which are damaged, incomplete or have loose parts.Return damaged products together with any loose parts for repair.Do not attempt to do any repairs yourself.Check the device and accessories for damaged, loose or missing parts, hygiene and completeness.Check all connection cables and tubes/hoses for damage.Check instruments and accessories for damage, sharp edges, loose or missing parts and rough surfaces.'check all connection tubes/hoses for damage, hygiene and completeness.'check o--rings and sealing caps as well as insulating parts for perfect condition.Important! in order to carry out the function checks, the devices must be in flawless condition and set up correctly.Carry out a visual check in order to guarantee this condition.Piranha suction pump: the piranha suction pump is equipped with automatic function checks which automatically detect and report their various errors or faults.The errors or faults detected may occur during the self--test or during operation.Switch on the power switch.'after each switch--of the power switch the device runs through an automatic function test (approx.10 seconds).Any fault or error detected is displayed on the lcd display (see section 5.7).'on successful completion of the test, the piranha suction pump is operational.Connect the complete system in accordance with section 3.Richard wolf medical instruments corporation (rwmic) considers this case closed.In the event that rwmic receives additional information a follow up report will be submitted to the fda.
|