BARD ACCESS SYSTEMS SITE-RITE 8 POWER SUPPLY SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Catalog Number 9770504
Device Problems Inappropriate/Inadequate Shock/Stimulation (1574); Sparking (2595)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard. The device has not been returned to the manufacturer, at this time, for evaluation. A lot history review (lhr) of dycvzz515 showed one other similar product complaint(s) from this lot number. The complaints for this lot number (dycvzz515) have been reported from same u. S. Facility. Device not returned at this time.
 
Event Description
Bd clinical specialist, reported that the equipment roll stand used with sr 8 ultrasound is shocking the users each time that they touch the stand. Users also mention that the plug causes a spark each time it is plugged into an outlet. A power strip is not used on the roll stand. First reported event for this power supply.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSITE-RITE 8 POWER SUPPLY
Type of DeviceSYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
DYMAX CORP.
141 zehner school road
zelienople PA 16063
Manufacturer Contact
shelly gilbert
605 n. 5600 w.
salt lake city, UT 84116
8015225640
MDR Report Key7936942
MDR Text Key122856471
Report Number3006260740-2018-02744
Device Sequence Number1
Product Code IYO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/04/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number9770504
Device Lot NumberDYCVZZ515
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location Hospital
Date Manufacturer Received09/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-