RESPIRONICS CALIFORNIA, INC V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Blower (738); Mechanical Problem (1384); Failure to Power Up (1476); Device Displays Incorrect Message (2591)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
Malfunction
|
Manufacturer Narrative
|
Event date: (b)(6) 2018.
Date of report: 9oct2018.
Customer swapped the power management (pm) board with another unit and it corrected the issue.
Customer swapped pm board back to other unit.
The customer replaced the pm board to resolve the reported issue.
No parts returned to failure investigation; therefore, the root cause of the reported issue could not be determined.
|
|
Event Description
|
Customer contacted technical support (ts) stating that unit did not power up and had error code messages of 35v supply failed, backup alarm failed, auxiliary alarm supply failed, and blower stalled.
The customer reported there was no patient involvement.
The event date was not specified; estimate used.
|
|
Search Alerts/Recalls
|
|
|