MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® SKYLITE¿ TIPLESS NITINOL STONE BASKET SKYLITE BASKET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® SKYLITE¿ TIPLESS NITINOL STONE BASKET SKYLITE BASKET Back to Search Results
Catalog Number 041900
Device Problems Loose or Intermittent Connection (1371); Difficult to Remove (1528); Physical Resistance/Sticking (4012)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The information provided by bard represents all of the known information at this time. Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the physician had to remove the handle of the skylite basket as it was stuck in the patients ureter with a stone. The physician wasn't able to properly remove the wire from the basket after loosening and removing the two black screws in the sheath of the basket. The wire ended up being cut in order to detach the handle. There were reports of a 5-10 minute surgical delay. No medical intervention was required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameBARD® SKYLITE¿ TIPLESS NITINOL STONE BASKET
Type of DeviceSKYLITE BASKET
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key7948186
MDR Text Key123255374
Report Number1018233-2018-04638
Device Sequence Number1
Product Code FFL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/13/2019
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number041900
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/26/2018
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received01/22/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-