• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110V; JCX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110V; JCX Back to Search Results
Catalog Number PMX110
Device Problem Failure to Pump (1502)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/11/2018
Event Type  malfunction  
Manufacturer Narrative
Results: one of the fuses was blown inside the pump max.The device was non-functional, therefore, functional testing was unable to be performed.The pump max did not power on.Exchanging the blown fuse with a new one allowed the device to power on momentarily before the new fuse blew out.Conclusions: evaluation of the returned pump max revealed a device with a blown fuse.The pump max was opened and the blown fuse was replaced.Subsequently powering on the pump max allowed the device to power on momentarily before the new fuse blew out.This was likely due to a short in the circuit.The root cause and location of the short were not able to be determined.Penumbra pumps are visually inspected and functionally test during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 110 (pump max).During the procedure, the pump max was unable to turn on.The procedure was therefore completed using a new pump max.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 110V
Type of Device
JCX
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key7949586
MDR Text Key123410692
Report Number3005168196-2018-01997
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00814548012773
UDI-Public00814548012773
Combination Product (y/n)Y
Reporter Country CodeCA
PMA/PMN Number
K122756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,09/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPMX110
Device Lot NumberF39387-17
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received09/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/13/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-