• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Failure to Power Up (1476)
Patient Problem No Patient Involvement (2645)
Event Date 09/12/2018
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement. The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601). Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). A livanova field service representative was dispatched to the facility to investigate. The service representative was able to confirm the reported issue and traced the failure to a defective mains plug. It was found that the contacts on the live and neutral connections in the mains plug were loose. The live connection had burnt back to a point where it lost contact. The technician shortened the cable and replaced the mains plug to resolve the reported issue. Subsequent functional verification testing was completed without further issues and the unit was returned to service. A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t was not powering on during in-service. There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key7951550
MDR Text Key124666656
Report Number9611109-2018-01289
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 10/10/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received10/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number16-02-80
Was Device Available for Evaluation? No Answer Provided
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received09/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/27/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-