Catalog Number 284002
Device Problem Pressure Problem (3012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The defect reported has been verified and repaired.Performed service and repair functions.Reviewed service history.Attach box label, software upgrade form, and fms vue final testing.Replaced springs on pressure arms with tip replacement kit to correct issue.Also replaced damage console cover, front bezel and membrane switch, and left side door assy.Performed software upgrade to the latest versions.The unit passed all diagnostic tests, functional tests, and is fully operational.Details can be found in the attached reports.A review of the device history record indicated that this batch of product was processed without incident; therefore, there is no evidence of manufacturing anomalies on the paperwork reviewed.A review into the depuy synthes mitek complaints system revealed no other complaints for this device's serial number.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).This report is being filed from the (b)(4).Complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
 
Event Description
It was reported by the customer that during a shoulder procedure, it was observed that the pump pressure rose up to 120 and went down to low pressure.The day set was replaced but the problem persisted.It was unknown how the procedure was completed and there were no patient consequences nor delays in the procedure.There was patient involvement reported.There were no injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FMS VUE PUMP
Type of Device
DISTENSION UNIT, FLUID, ARTHROSCOPIC 
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
jennifer lawrence
325 paramount drive
raynham, MA 02767
5088808100
MDR Report Key7955867
MDR Text Key123505264
Report Number1221934-2018-54921
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705020591
UDI-Public10886705020591
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number284002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2015
Date Manufacturer Received10/29/2015
Date Device Manufactured07/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-