Model Number PB1018 |
Device Problems
Fracture (1260); Gradient Increase (1270)
|
Patient Problems
Pulmonary Valve Stenosis (2024); No Consequences Or Impact To Patient (2199)
|
Event Date 09/20/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Product analysis: the product remains implanted therefore no product analysis can be completed.Conclusion: without return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that an unknown time following the implant of this transcatheter pulmonary bioprosthetic valve, the valve was reported to have multiple stent fractures.A second non-medtronic valve was implanted inside of the valve.No additional adverse patient effects were reported. .
|
|
Manufacturer Narrative
|
Additional information received that approximately five years and nine months post implant of the pulmonary valve, the valve was replaced due to elevated pressures and gradients with severe stenosis.A valvuloplasty was performed followed by stent placement and an additional dilatation.A non-medtronic valve was then successfully implanted.No additional adverse patient effects were reported. if information is provided in the future, a supplemental report will be issued.
|
|