• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY SR RF; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY SR RF; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM1240
Device Problems Pacemaker Found in Back-Up Mode (1440); Operating System Version or Upgrade Problem (2997)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/27/2018
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented during a follow up in clinic.The pacemaker went into backup vvi mode upon a failed attempt to perform cyber security update in clinic.Device restoration was not successful.No intervention was performed at the time.Additional information was not reported.
 
Event Description
New information received on 10/22/2018 indicated that patient presented for procedure.The device was explanted and replaced.The patient was stable post procedure.
 
Manufacturer Narrative
The reported field event of backup vvi (bvvi) was confirmed in the laboratory.The device was tested on the bench the cause of the bvvi was consistent with hardware anomaly.Correction: this supplemental report is to correct the previously submitted report to include patient ethnicity and race.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY SR RF
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7958757
MDR Text Key123501656
Report Number2017865-2018-15215
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734507066
UDI-Public05414734507066
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/11/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2016
Device Model NumberPM1240
Device Lot NumberA000007922
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/17/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-