Model Number PM1240 |
Device Problems
Pacemaker Found in Back-Up Mode (1440); Operating System Version or Upgrade Problem (2997)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/27/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that the patient presented during a follow up in clinic.The pacemaker went into backup vvi mode upon a failed attempt to perform cyber security update in clinic.Device restoration was not successful.No intervention was performed at the time.Additional information was not reported.
|
|
Event Description
|
New information received on 10/22/2018 indicated that patient presented for procedure.The device was explanted and replaced.The patient was stable post procedure.
|
|
Manufacturer Narrative
|
The reported field event of backup vvi (bvvi) was confirmed in the laboratory.The device was tested on the bench the cause of the bvvi was consistent with hardware anomaly.Correction: this supplemental report is to correct the previously submitted report to include patient ethnicity and race.
|
|