TERUMO CARDIOVASCULAR SYSTEMS CORPORATION SARNS 8000 UNIVERSAL ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER
|
Back to Search Results |
|
Model Number 16402 |
Device Problem
Break (1069)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/24/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The reported issue was discovered during student setup training.As per the field service representative (fsr), the occlusion knob was very hard to turn, which had probably led to the broken tongue thus knob was cleaned and lubricated.He replaced the tongue.The suspect device will be returned to the manufacturer for further evaluation.
|
|
Event Description
|
It was reported that during the use of the device for a non-clinical activity, the molded tongue of the pump was broken.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|