Correction: three devices; not one device, were reported to have disconnected.Date returned to mfr - one device was returned on 10/18/2018 and the other two were returned on 10/22/2018.Three devices were received for evaluation.During visual inspection, the tubing was observed separated from the 2-piece luer locks in the samples.There are trace of solvent at tubing end where separation observed.All samples were tested with 8psi under water pressure test and no leaks were observed.The reported condition was verified.The cause was due to a machine in the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|