Model Number SYRINGES |
Device Problems
Leak/Splash (1354); Manufacturing, Packaging or Shipping Problem (2975)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/30/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product not yet returned for evaluation.Refer customer to the distributor for replacement.Most likely underlying root cause: mlc-140- supplier manufacturing defect.Based on the post market review (b)(4) and review of the fda's total product life cycle (tplc) - total device problems, for possible delay in administration of insulin.Reference - capa (b)(4).
|
|
Event Description
|
Pharmacist is calling on behalf of the customer.Pharmacist says that customer brought back in two boxes of true plus syringes.Customer says that syringes are not drawing up insulin and when it does, insulin leaks during injection.Pharmacy wants replacement syringes.Advised customer to contact distributor for replacement syringes.Advised that i will send return envelope for some of the syringes for investigation purposes.
|
|