Model Number SYRINGES
Device Problems Leak/Splash (1354); Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/30/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product not yet returned for evaluation.Refer customer to the distributor for replacement.Most likely underlying root cause: mlc-140- supplier manufacturing defect.Based on the post market review (b)(4) and review of the fda's total product life cycle (tplc) - total device problems, for possible delay in administration of insulin.Reference - capa (b)(4).
 
Event Description
Pharmacist is calling on behalf of the customer.Pharmacist says that customer brought back in two boxes of true plus syringes.Customer says that syringes are not drawing up insulin and when it does, insulin leaks during injection.Pharmacy wants replacement syringes.Advised customer to contact distributor for replacement syringes.Advised that i will send return envelope for some of the syringes for investigation purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGES
Type of Device
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
karen devincent
2400 nw 55th court
fort lauderdale, FL 33309
954677-920
MDR Report Key7973565
MDR Text Key123966767
Report Number1000113657-2018-01050
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
PMA/PMN Number
K993017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 10/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2018
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberSYRINGES
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Date Manufacturer Received10/04/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SECOND THERAPY
-
-