Information was received indicating that while the nurse was attempting to pull out a smiths medical jelco® cathlon® i.V.Catheter, the plastic part broke off in the patient.It was reported that approximately 2mm of the plastic came out.Subsequently, the patient had the severed catheter removed in interventional radiology.There were no further reported adverse patient effects.
|
Device evaluation summary: one jelco cathlon iv catheter assembly was returned for evaluation.Immediate inspection of the sample showed the catheter had been cut off at approximately 5.9mm from the hub nose.Further examination of the cut showed the surface was jagged and highly damaged.During production, to assure that the products meet functional requirements the dimensions of the catheter components (eyelet, tubing and hub) are tightly controlled.Once assembled, these products are 100% inspected to demonstrate the catheter tube is properly secured to the hub.In addition, catheter samples from each lot number are subjected to destructive testing to determine minimum break strength.Based on the investigation, the complaint allegation was confirmed.User interface is noted as the problem source, due to evidence that the damage of the device was consistent with the use of the catheter against the device ifu (instructions for use); either through a needle reinsertion or use of a lancet or scissor near the catheter.
|