Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the defibrillation, pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
Boston scientific received information that the non-boston scientific right ventricular (rv) lead connected to this implantable cardioverter defibrillator (icd) experienced many episodes of noise with oversensing.The patient was shocked inappropriately due to oversensing of the rv lead noise.It was reported that there was not pacing inhibition and that the patient had an underlying rhythm of 71 beats per minute.Sensing, impedance, and thresholds all appeared to be normal.Rv lead fracture was suspected, but was not able to be confirmed via x-ray.The non-boston scientific rv lead was surgically abandoned and the icd was explanted and replaced.No additional adverse patient effects were reported.
|