Model Number MODEL 3000
Device Problems Computer Software Problem (1112); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/26/2018
Event Type  malfunction  
Manufacturer Narrative
Suspect device software version: version 1.5.2.1.
 
Event Description
It was reported that high impedance was observed on the patient's m102 vns using a vns programming tablet with m3000 v1.5.2.1 software.However, upon running diagnostics after changing the vns output current to 0.00 ma, there was no high impedance present.There was no resulting treatment decision affected by this event.This issue was duplicated with in-house testing.For model 102/102r generators programmed to output currents greater than 1.00 ma , it was observed that system diagnostics using m3000 v1.0 and m250 software would display ok lead impedance while system diagnostics using m3000 v1.5 software would display a false high impedance.It was communicated to the patient¿s physician that the high impedance observed on the m3000 v1.5.1.2 software was false.It was determined that the cause of this false high impedance message was a software error on m3000 v1.5.1.2 software; when system diagnostics were performed by the user with m3000 v1.5.2.1 software on m102/102r generators (normal mode output current >0 ma), normal mode diagnostics would actually be performed instead.The execution of normal diagnostics instead of system diagnostics is not apparent to the user, and the returned dcdc code was being compared against system diagnostic thresholds.It was determined that m102/102r system diagnostics functioned as expected when the generator was programmed to 0 ma and provided a true impedance value.No additional relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGRAMMING SOFTWARE
Type of Device
PROGRAMMING SOFTWARE
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key7980531
MDR Text Key125772873
Report Number1644487-2018-01863
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 10/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 3000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received09/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/15/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age71 YR
-
-